Dynamic Strain Measuring Instrument DAS-406C

Model DAS-406C / Minebea Mitsumi Inc. (Official website)

DC amplification type dynamic strain measuring instrument. Frequency response from DC to 200 kHz.

Last verified: (date the content was verified against the source document)

Specifications

Applicable gauge resistance 60 to 1k Ω
Frequency response range DC to 200 kHz
Non-linearity 0.01 %F.S.
Power supply voltage AC100 V±10 V
Mass 1.5 kg

How to choose Strain measuring instrument

Points to check when selecting, with the values for this product (DAS-406C).

  1. 1. Choose by static or dynamic, based on frequency bandwidth

    This product: Frequency response range DC to 200 kHz

    Check that the frequency response range is sufficient for the speed of change of the phenomenon to be measured. For mainly static loads or temperature changes, a static strain measuring instrument is suitable; for vibration or shock, a dynamic strain measuring instrument is suitable. Some DC amplifier types handle high frequencies such as DC to 200 kHz, while carrier amplifier types are resistant to noise but have a comparatively narrower bandwidth.

  2. 2. Confirm compatibility with the gauge resistance

    This product: Applicable gauge resistance 60 to 1k Ω

    Strain gauges are commonly 120 Ω or 350 Ω, so check that the resistance is within the connectable range of the instrument. Outside that range the bridge cannot be balanced and measurement is not possible. Also check whether the gauge factor setting range allows entering the value of the gauge used.

  3. 3. Estimate accuracy from measuring range and non-linearity

    This product: Non-linearity 0.01 %F.S.

    Confirm that the maximum expected strain falls within the measuring range, then estimate the error from non-linearity (0.01 to 0.1 %F.S. for the listed products). Since %F.S. is based on full scale, measuring small strains over a wide range increases the relative error.

  4. 4. Balance bridge voltage against heat generation and sensitivity

    A higher bridge voltage increases output and improves the S/N ratio, but it also increases drift caused by self-heating of the gauge. Check whether a voltage suitable for the thermal conductivity of the bonded material, the gauge dimensions, and the recommended excitation voltage of the transducer can be selected.

  5. 5. Confirm the required number of simultaneous measurement points

    When acquiring stress distribution at multiple points simultaneously, insufficient channels prevent obtaining time-aligned data. Check whether the system can be expanded with add-on units, and whether each channel has an independent amplifier.

  6. 6. Long cables and ease of use in the field

    Long cables reduce sensitivity due to resistance, so check whether there is a lead wire length correction function. Also select according to the recording system and site conditions: wire break check, portable or benchtop type, output format, communication such as RS-232C, and power supply type.

This guide was written by AI based on the listed products' documents and specification items.

About Strain measuring instrument and all products

Description

Equipped with a 3 1/2 digit digital monitor, allowing setting of an arbitrary standard equivalent strain.

Features

  • DC to 200 kHz
  • 3 1/2 digit digital monitor

Applications

  • Dynamic strain measurement

FAQ

Q. What frequency range of dynamic strain can the DAS-406C measure?
A. The DAS-406C is a DC amplification type dynamic strain measuring instrument with a frequency response from DC to 200 kHz. The frequency response range is DC to 200 kHz.
Q. What is the range of strain gauge resistance that can be used with the DAS-406C?
A. The applicable strain gauge resistance of the DAS-406C is 60 to 1k Ω.
Q. What are the non-linearity, power supply voltage, and mass of the DAS-406C?
A. The DAS-406C has a non-linearity of 0.01 %F.S., a power supply voltage of AC100 V±10 V, and a mass of 1.5 kg.
Q. What can you tell me about the display and setting functions of the DAS-406C?
A. The DAS-406C is equipped with a 3 1/2 digit digital monitor and allows setting of an arbitrary standard equivalent strain.